MEDOS INTERNATIONAL SARL CERENOVUS ENTERPRISE; INTRACRANIAL NEUROVASCULAR STENT Back to Search Results
Catalog Number ENCR402312
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer ref#: (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section h3: the device is available to be returned for evaluation and testing.However, it has not been received to date.If the device returns, a device investigation will be performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
A healthcare professional reported that during an angioplasty, an enterprise2 4mmx23mm vascular reconstruction device (encr402312, 9477853) was impeded in the proximal end of a prowler select plus 150/5cm microcatheter (606s255x, lot unknown) and could not advance anymore.The doctor removed the stent and microcatheter (mc) from the patient and switched to a new stent to complete the procedure with the same microcatheter (mc).No additional information is available.
 
Additional Manufacturer Narrative
Manufacturer ref# (b)(4).Updated sections on this medwatch: b4, d9, g3, g6, h2, h3, h6 and h11.Complaint conclusion: a healthcare professional reported that during an angioplasty, an enterprise2 4mmx23mm vascular reconstruction device (encr402312, 9477853) was impeded in the proximal end of a prowler select plus 150/5cm microcatheter (606s255x, lot unknown) and could not advance anymore.The doctor removed the stent and microcatheter (mc) from the patient and switched to a new stent to complete the procedure with the same microcatheter (mc).No additional information is available.The device was returned to j&j medtech for further evaluation.A non-sterile enterprise2 4mmx23mm vascular reconstruction device was received contained in the decontamination pouch.Upon receiving the device, visual inspection was performed, and no appearance of damage was observed.A functional test was performed.A lab sample of the prowler select plus was flushed using a lab syringe.The unit was then inserted into prowler select plus, and it advanced smoothly without any resistance or friction as the stent passed through.The stent was successfully expelled from the distal tip of the microcatheter.A device history record (dhr) was performed and it indicated this product was final inspection tested at lake region medical and was determined to be acceptable.As the functional test was performed without issues, the impeded condition encountered during the procedure could not be confirmed.However, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the analysis.As part of the j&j medtech quality process, all devices are manufactured, inspected, and released to approved specifications.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no capa activity is required at this time.It should be noted that multiple factors could cause product failure.The instructions for use (ifu) do contain the following recommendations: the introducer must be properly engaged with the infusion catheter hub to enable stent introduction into the infusion catheter.Do not apply undue force if resistance is encountered at any point during stent manipulation.Withdraw the unit and advance to a new one.If resistance is felt while recapturing the stent, do not continue to recapture the device.Withdraw the infusion catheter slightly to unsheathe the stent (without exceeding the recapture limit), and then attempt to recapture the stent again.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERENOVUS ENTERPRISE
Common Device Name
INTRACRANIAL NEUROVASCULAR STENT
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC
325 paramount dr
raynham MA 02767
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
9497882078
MDR Report Key22385693
Report Number3008114965-2025-00668
Device Sequence Number18410895
Product Code NJE
UDI-Device Identifier10886704075356
UDI-Public(01)10886704075356(17)271128(10)9477853
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
H60001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberENCR402312
Device Lot Number9477853
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2025
Supplement Date Received by Manufacturer07/29/2025
Initial Report FDA Received Date07/02/2025
Supplement Report FDA Received Date08/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/16/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PROWLER SELECT PLUS 150/5CM.
Patient SexUnknown
-
-