There was no device available for analysis; therefore, no physical or visual analysis of the product could be performed.The device history record (dhr) review was unable to be performed as the lot number was not provided.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the events of disfigurement, pain, length too long were defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.The reported patient symptoms are known risks associated with this device type and indicated as such in the instructions for use.The reported issue with length too long may have been due to a potential measurement error during the original implantation procedure.Based on the information available, the conclusion codes of unintended use error caused or contributed to event and known inherent risk of device were assigned to this investigation.Correction to field b1: adverse event/product problem
detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.
|