|
Additional information: d9, g3, h3, h6.- one unpackaged, sabre, sj, 3.0 mm x 7 cm, ar-7300sr, batch 15046758, was received for investigation.- visual evaluation, under a magnifier, noted damage on the edges of the outer tip.Abrasion marks were also evident on the inner tube.- functional testing noted that the device functioned as intended without producing debris.-the most likely cause for the observed condition is the application of excessive forces/side-loading on the device during use.- dfu-0215-eo; f-7 indicates that "excessive side-loading" can thus cause irreversible damage to the device and produce metal shavings.-refer to investigation photos.- the complaint allegation that the device is damaged is confirmed.
|