ARTHREX, INC. DRIVER SHAFT, T25 HEXALOBE, CANNULATED; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number DRIVER SHAFT, T25 HEXALOBE, CANNULATED
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 06/06/2025, it was reported by a sales representative via (b)(4) that an ar-8610d-65 drive shaft broke during insertion.Issue discovered during a case with no patient effect.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Complaint allegation is not confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause of this failure is attributed to wear and tear damage incurred over repeatedly torquing/engaging the driver within the screw head and extended use in the field.Manufacturing date 2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVER SHAFT, T25 HEXALOBE, CANNULATED
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22386285
Report Number1220246-2025-02716
Device Sequence Number15238371
Product Code LXH
UDI-Device Identifier00888867197299
UDI-Public(01)00888867197299(11)191015(10)1391939
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 09/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDRIVER SHAFT, T25 HEXALOBE, CANNULATED
Device Catalogue NumberAR-8610D-65
Device Lot Number1391939
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date06/06/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2025
Supplement Date Received by Manufacturer09/26/2025
Initial Report FDA Received Date07/02/2025
Supplement Report FDA Received Date09/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2019
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-