• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PRLNE BLU 4X24IN 5-0 D/A C-1; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PRLNE BLU 4X24IN 5-0 D/A C-1; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Catalog Number M8725
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.It was reported when they opened the sutures, there's m8726 inside the boxes sutures.Not the correct ones.No adverse impact patient/user.Device is available for return.No additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 7/2/2025.H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 8/1/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.Additional information: d9, h3, h6.H3 investigational summary: the product was returned to ethicon for evaluation.Three representative unopened samples that pertain to product code m8725, lot 103urb; one representative unopened sample that pertains to product code m8726, lot 105mh8 and five representative unopened samples that pertain to product code m8726, lot 106ba3 were received for analysis.Due to this mismatch, a further investigation was performed.The lot numbers were reviewed against our system and revelated that the product code m8725, lot 103urb was packaged on october 8, 2024.The product m8726, lot 105mh8 was packaged on january 10, 2025.The product m8726, lot 106ba3, was packaged on february 10, 2025.According to product code m8725, lot 103urb, there was a difference of three to four months between the manufacturing dates of the products.Therefore, these lots were not mixed at our manufacturing site.Additionally, the box was not received.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what caused the reported complaint.It should be noted that as part of our quality process, the device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE BLU 4X24IN 5-0 D/A C-1
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key22386955
Report Number2210968-2025-07802
Device Sequence Number7100823
Product Code GAW
UDI-Device Identifier10705031045910
UDI-Public(01)10705031045910(17)290930(10)103URB
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberM8725
Device Lot Number103URB
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/11/2025
Supplement Date Received by Manufacturer07/08/2025
Initial Report FDA Received Date07/02/2025
Supplement Report FDA Received Date08/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-