• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG EVIA DR-T; PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG EVIA DR-T; PACEMAKER Back to Search Results
Model Number 359529
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pocket Erosion (2013)
Date of Event 06/18/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Erosion was observed following the implantation of this biotronik device.The sterilization process was investigated.The validated process assures that all sterilization parameters, such as gas concentration, temperature, humidity, etc., are within its specified ranges for each distributed device.Additionally, an analysis of validated microbiological indicators is performed after every sterilization procedure as evidence of successful completion of the sterilization process.Review of the biotronik complaint database did not reveal any changes regarding the trend for this type of incident.In summary, the erosion was not device related.
 
Event or Problem Description
This device was explanted due to device pocket was beginning to show signs of erosion.Should additional information become available, this file will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIA DR-T
Common Device Name
PACEMAKER
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
DE  12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key22388983
Report Number1028232-2025-03180
Device Sequence Number7053743
Product Code NVZ
UDI-Device Identifier04035479101312
UDI-Public(01)04035479101312(17)150331
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P950037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date03/31/2015
Device Model Number359529
Device Catalogue Number359529
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2025
Initial Report FDA Received Date07/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/09/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age89 YR
Patient SexMale
-
-