Catalog Number PR-35041-HPHNM
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "catheter kinking and breakage issue reported from cvad department.The department have shared images of 6-month-old catheter with the mentioned complaint.Anatomical insertion site of device: peripherally upper arm catheter flow got stuck, technicians have taken back the broken catheter from body.There was a delay in the procedure." 3 devices were reported.Associated mdrs: 9680794-2025-00542 and 9680794-2025-00540.
 
Event or Problem Description
It was reported that "catheter kinking and breakage issue reported from cvad department.The department have shared images of 6 month old catheter with the mentioned complaint.Anatomical insertion site of device: peripherally upper arm catheter flow got stuck, technicians have taken back the broken catheter from body.There was a delay in the procedure." 3 devices were reported.Associated mdrs: 9680794-2025-00542, 9680794-2025-00540 and 9680794-2025-00541.
 
Additional Manufacturer Narrative
(b)(4) the actual device was not returned; however, the customer provided six photos for analysis.The first two images show two different pressure injectable (pi) catheters whilst inside the patient.The remaining four images show kinking on the catheter body.A rupture on the catheter body was also observed.The customer was contacted to address the rupture.No clarification was provided after three attempts.A rupture of this type can be a result of kinking along the catheter body in combination with high pressurization; however, as no clarification was provided, the circumstances cannot be verified.Despite the damage observed in the customer images, a complete visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.The ifu provided with the kit informs the user, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained and further consultation requested".The ifu also states, "use only lumen(s) labeled 'pressure injectable' for pressure injection to reduce risk of catheter failure and/or patient complications".The ifu also states, "ensure catheter patency prior to use, including prior to pressure injection.Do not use syringes smaller than 10 ml to reduce risk of intraluminal leakage or catheter rupture.Power injector equipment may not prevent over pressurizing an occluded or partially occluded catheter".The customer report of kinked catheter was confirmed by visual inspection of the customer supplied photo.Analysis of the photos also revealed evidence of a catheter body rupture.Rupturing of this type can be a result of kinking in combination with high pressurization.Regardless of the observed damage, a full complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW PI PICC 1L 4FRX50CM W/80CM HYDRO NIT SWG
Common Device Name
CATHETER, INTRAVASCULAR, THER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key22390764
Report Number9680794-2025-00541
Device Sequence Number9104599
Product Code LJS
UDI-Device Identifier00801902101051
UDI-Public(01)00801902101051(11)240327(17)260228(10)33F24B0693(21)
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K113277
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Catalogue NumberPR-35041-HPHNM
Device Lot Number33F24B0693
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2025
Supplement Date Received by Manufacturer07/25/2025
Initial Report FDA Received Date07/02/2025
Supplement Report FDA Received Date07/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED.; NONE REPORTED.
Patient SexMale
-
-