ARTHROCARE CORPORATION ULTRABUTTON ADJUSTABLE FIXATION DEVICE; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72290003 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Erythema (1840); Fluid Discharge (2686); Swelling/ Edema (4577)
|
| Date of Event |
05/02/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication reported that includes reference to the use of a smith+nephew product without evidence about a specific product problem.The reported complication relates to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, after a surgical procedure on (b)(6) 2025, on a patient presenting acromioclavicular joint dislocation, in which an ultrabutton fixation device was used, the wound healed well and the patient was discharged.After discharge, the wound showed redness, swelling, and abnormal exudation.Therefore, the patient was readmitted on (b)(6) 2025 and underwent debridement and suturing surgery.The wound healed well after the surgery without any obvious complications.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h2: additional information on a2, a3, e1, e2 & e3.
|
| |
|