MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG; AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND
|
Back to Search Results |
|
| Model Number |
MMT-1715K |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted as part of remediation activities associated with capa # 643143, which is addressing mdr reporting guidance from fdas 1995 preamble, which ensures complaints related to corrections and recalls which were previously reported to fda under 21 cfr 806 are now reported as mdrs; this is not a new malfunction/event.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Pump was received with high sleep current due to moisture damaged electronic assembly on pcb1.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported to medtronic minimed that the customer experienced replace battery now alarms without prior low battery alerts, with the battery depleting rapidly after being replaced.The customer reported no adverse event.The event involved product(s) mmt-1715k.Troubleshooting was performed which included reviewing alarm history, confirming the use of new alkaline batteries, and verifying that the battery cap was not loose, cracked, or damaged.The issue persisted, and the pump was replaced.The customer was advised to monitor the insulin pump and continue using it until the replacement arrived.No harm requiring medical intervention was reported.The product was returned for analysis.
|
| |
|
Search Alerts/Recalls
|
|
|