| Model Number |
VLFT10GEN |
| Medical Device Problem Codes |
Energy Output Problem (1431); Device Displays Incorrect Message (2591); Connection Problem (2900)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
06/09/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a laparoscopic gastric sleeve procedure, the device had an e402 error code right after activation.Repeated firm grasping/activating would get an end-tone and evidence of energy delivery when applied to the normal thickness of short gastric vessels.The jaws of the device were cleaned frequently.Patient bled after cutting and needed blood transfusion.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: lf5637, ligasure lf5637 retractable l hook (lot#:50570263x).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a laparoscopic gastric sleeve procedure; the device had an e402 error code right after activation.Repeated firm grasping/activating would get an end-tone and evidence of energy delivery when applied to the normal thickness of short gastric vessels.The jaws of the device were cleaned frequently.Procedure was completed using a sonicision device.Patient had 250 ml of blood loss after cutting and needed blood transfusion.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: a2 (date of birth), b2, b5, d3 (mfr name and street 1), g3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|