CHRISTOPH MIETHKE GMBH & CO KG M.BLUE PLUS VALVE; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX804T
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 06/05/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection: during the investigation, scratches on the outer housing of the valves, but no significant deformations were determined.Permeability test: a permeability test has shown that both valves are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical position.The results show that the valves are operating within the specified tolerances in their respective relevant position.Adjustment test: the valves were tested and both valves are adjustable in all pressure settings.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected and the braking force is within the given tolerance.Internal inspection: after dismantling of the valves, organic deposits were found in both valves.Results: based on our investigation results, we can determine visible deposits in both valves.The deposits had no effect upon the specifications at the time of the examination.The shuntsystem operated within all specifications.Deposits caused by substances naturally present in the patient's body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic material can affect the integrity of the valve.The examination of the return has been completed with the drafting of this report.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that an m.Blue plus (#fx804t) was implanted during a procedure performed on an unspecified date in 2020.According to the complainant, the shunt system caused an underdrainage.The patient underwent a revision procedure on (b)(6) 2025.No patient complications were reported as a result of the revision procedure.The complained device was returned to the manufacturer for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE PLUS VALVE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key22394159
Report Number3004721439-2025-00176
Device Sequence Number9061050
Product Code JXG
UDI-Device Identifier04041906504793
UDI-Public4041906504793
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K192266
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Expiration Date03/25/2024
Device Model NumberFX804T
Device Catalogue NumberFX804T
Device Lot Number20041317
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/18/2025
Initial Report FDA Received Date07/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/27/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age32 YR
Patient SexMale
Patient Weight80 KG
-
-