MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 REACT CATHETER; CATHETER, PERCUTANEOUS, NEUROVASCULATURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 REACT CATHETER; CATHETER, PERCUTANEOUS, NEUROVASCULATURE Back to Search Results
Model Number REACT-71
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received a report that a solitaire encountered resistance in the middle of the react and the stent pushwire was broken in the distal section.The patient was undergoing emergency interventional thrombectomy for treatment of ischemic stroke in the middle cerebral artery.Patient condition: mrs baseline: 4, mrs procedure: 4, nihss baseline: 15, nihss post procedure: 4, tici baseline: 0, tici post procedure: 3.Iv tpa was not contraindicated.There was 1 pass with the device.It was noted the patient's vessel tortuosity was normal.The access vessel was the femoral artery with an 11mm diameter.Patient has a medical history of hypertension level 2.Stroke onset to reperfusion time was 4.5 hours.It was reported that on (b)(6) 2025, the surgeon was performing an emergency thrombectomy on a patient with acute large artery occlusion.The lesion was located at the middle cerebral artery at the end of the internal carotid artery.The surgeon planned to use the solitact technology to remove the thrombus.After the micro-guidewire and micro-catheter established the pathway, the micro-catheter performed proximal and distal angiography to determine the proximal and distal positions of the thrombus removal and prepare for the delivery of the thrombectomy stent.The stent selected was sfr-6-30.The surgeon reported that the delivery and deployment were smooth.After staying for five minutes, the react-71 was pushed to go upper, and carry out the combination of drawing and pulling.After the stent was withdrawn, it was found that the thrombectomy stent was broken in the react-71.The stent and react-71 were replaced, and the thrombus was removed for the second time.The surgery was successfully completed.The pushwire was not torqued during the procedure.Resistance was encountered.All broken segments of the pushwire were removed from the patient.Since the segments were broken in the suction catheter, the surgeon removed the suction catheter and changed the access.The stent was removed from the patient.The reported devices and any accessory devices were prepared and the catheter was flushed as indicated in the instructions for use (ifu).No patient symptoms or complications were associated with this event.Ancillary devices include a ced medical-27 microcatheter.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information received reported that the cause of the resistance and pushwire break was not determined.The catheter was vi sually inspected and there was no problem, and the stent was inside the catheter.The tip of the solitaire sheath was firmly secured in the hub of the microcatheter.
 
Additional Manufacturer Narrative
Product analysis: as found condition: the solitaire fr 6-30 stent was returned for analysis, stuck inside the returned react-71 catheter, enclosed in a sealed biohazard bag and a shipping box.The solitaire stent pusher wire was not returned.The reported ced medical 27 microcatheter was not returned.Additionally, the model was not reported.Damaged location details: the solitaire fr 6-30 stent¿s working and non-working length struts were in good condition.The glue domes outside the proximal marker were observed to be damaged.It was observed that the solitaire stent was broken at the detachment zone.However, the proximal broken end of the pusher wire was not returned.Therefore, no sample could be sent out for scanning electron microscopy (sem) failure analysis testing.No voids or flashes were found on the react-71 catheter hub.No bends or kinks were found on the catheter body.The catheter distal tip/marker was found to be compressed/flattened.No other anomalies were found.Testing/analysis: the react-71 catheter was dissected (cut), and the solitaire fr 6-30 stent was removed.The total and usable lengths of the react-71 catheter were measured within the specified limits.The react-71 catheter was flushed with water and was found patent, tested via an in-house 0.050¿ mandrel through the hub without issues; however, resistance was observed at the distal tip.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REACT CATHETER
Common Device Name
CATHETER, PERCUTANEOUS, NEUROVASCULATURE
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
7635265677
MDR Report Key22394421
Report Number2029214-2025-01500
Device Sequence Number18398202
Product Code QJP
UDI-Device Identifier00847536044197
UDI-Public(01)00847536044197(17)270212(10)D019098
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K182097
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberREACT-71
Device Catalogue NumberREACT-71
Device Lot NumberD019098
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2025
Supplement Date Received by Manufacturer07/04/2025
08/28/2025
Initial Report FDA Received Date07/03/2025
Supplement Report FDA Received Date08/01/2025
08/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
Patient Weight65 KG
-
-