U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPTA QUAD CRT-P MRI SURESCAN; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPTA QUAD CRT-P MRI SURESCAN; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number W4TR01
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bacterial Infection (1735); Bacteremia (4846)
Date of Event 06/12/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the cardiac resynchronization therapy pacemaker (crt-p) system was explanted due to methicillin-sensitive staphylococcus aureus and bacteremia infections and replaced with a leadless implantable pulse generator (ipg).No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: 439888 lead implanted (b)(6) 2017.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Analysis of the returned device was inconclusive.Hybrid analysis revealed that there was no evidence of a high current drain condition on the hybrid and no hybrid anomalies were observed.Battery analysis revealed that no evidence of an internal mechanism for premature battery depletion was found during destructive analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEPTA QUAD CRT-P MRI SURESCAN
Common Device Name
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key22395065
Report Number3004209178-2025-11383
Device Sequence Number15743305
Product Code NKE
UDI-Device Identifier00643169735644
UDI-Public(01)00643169735644(17)181128
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P010015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date11/28/2018
Device Model NumberW4TR01
Device Catalogue NumberW4TR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2025
Supplement Date Received by Manufacturer03/20/2026
Initial Report FDA Received Date07/03/2025
Supplement Report FDA Received Date04/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-58 LEAD, 5076-52 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age76 YR
Patient SexFemale
Patient Weight104 KG
-
-