| Catalog Number |
136523000 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d6, h4: the information provided regarding the manufacturing date and implantation date are identified to be conflicting.Continued follow-up is being conducted and any additional information received will be reported.
|
| |
|
Event or Problem Description
|
|
It was reported that patient underwent initial surgery on an unknown date in 2019 to have a hip joint installed.After awhile, patient was admitted to the hospital with a hip dislocation in an extreme position.During revision surgery, it was revealed that the insert and the head were identically erased.Implant was removed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees hat the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: b5 and d10 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received states that during the revision surgery, it was discovered that the liner and head had been worn down.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.According to the information received, "hip replacement, revision patient have a hip joint installed, after a while patient was admitted to the hospital with an hip dislocation in an extreme position, during revision surgery it was revealed that the insert and the head were identically erased".A manufacturing record evaluation were performed for the finished device.It was manufactured on 16-oct-2018.Total 36 parts were manufactured per specification and all raw materials met specification.After searching "136523000/d18100383" in etq, no record is displayed.After searching the product description in nr list exported from the etq system, all results have been reviewed, no case even similar to the issue described in this complaint is identified.After searching " 136523000/d18100383" in sap by zqmr1023, no internal lock record is displayed.Hereby drawing a conclusion that no non-conformances were identified for this batch of products as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: a manufacturing record evaluation were performed for the finished device.It was manufactured on 16-oct-2018.Total 36 parts were manufactured per specification and all raw materials met specification.After searching "136523000/d18100383" in etq, no record is displayed.After searching the product description in nr list exported from the etq system, all results have been reviewed, no case even similar to the issue described in this complaint is identified.After searching " d18100383" in sap by zqmr1023, no internal lock record is displayed.Hereby drawing a conclusion that no non-conformances were identified for this batch of products device history review: a manufacturing record evaluation was performed for the finished device [136523000/d18100383] number, and no non-conformances / manufacturing irregularities were identified.
|
| |
|
Search Alerts/Recalls
|