• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3507235MC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 06/12/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: silent rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
A female patient who underwent breast augmentation revision with mentor memorygel, 235cc silicone prosthesis experience left sided silent rupture post procedure.The rupture was discovered during the surgery.As a result, the patient underwent bilateral replacements as follows: right (b)(6), left sn (b)(6) on (b)(6) 2025.
 
Additional Manufacturer Narrative
Suspect device received on july 22, 2025.Device evaluation completed on july 27, 2025: per additional information received with the device the patient also experienced right sided gel bleed.Since no serial number was available on the returned devices, it was not possible to determine which device corresponded to the left-side breast implant, therefore, both products were analyzed.The product was returned to mentor for evaluation.Mentor conducted visual inspection and leak testing of the returned device.Visual analysis of the returned device revealed that no damage or anomalies were observed on the breast implant.Leak testing was performed, in accordance with mentor procedures, and an area of shell abrasion was observed on the posterior view.In addition, a tear was noted within the shell abrasion measuring approximately 0.1 cm.These findings are consistent with normal wear of the device.Shell abrasion suggests in-vivo folding or creasing of the device, which may be caused by continuous and sustained stresses to the device, such as an excessively small breast pocket or the folding or wrinkling of the shell within the breast pocket.In some cases, breast implants may also experience normal wear over time.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Updated device evaluation completed on august 08, 2025: the product was returned to mentor for evaluation.Mentor conducted visual inspection and leak testing of the returned device.Visual analysis of the returned device revealed that no damage or anomalies were observed on the breast implant.Leak testing was performed, in accordance with mentor procedures, and an area of shell abrasion was observed on the posterior view.In addition, a tear was noted within the shell abrasion measuring approximately 0.1 cm.These findings are consistent with normal wear of the device.Shell abrasion suggests in-vivo folding or creasing of the device, which may be caused by continuous and sustained stresses to the device, such as an excessively small breast pocket or the folding or wrinkling of the shell within the breast pocket.In some cases, breast implants may also experience normal wear over time.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
9497882078
MDR Report Key22397827
Report Number1645337-2025-07360
Device Sequence Number20992339
Product Code FTR
UDI-Device Identifier00081317004110
UDI-Public(01)00081317004110(11)140912(17)190911(10)6862492
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/11/2019
Device Catalogue Number3507235MC
Device Lot Number6862492
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/12/2025
Supplement Date Received by Manufacturer07/22/2025
08/08/2025
Initial Report FDA Received Date07/03/2025
Supplement Report FDA Received Date07/30/2025
08/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/12/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-