H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Each event reported to bd is evaluated and investigated in accordance with our complaint investigation procedures.The investigation process includes, but is not limited to, evaluation of the event details provided by the complaint facility, a review of complaint history, manufacturing records and risk document where applicable, and an evaluation of the subject device when available to identify potential contributing factors.The device has not been returned to the manufacturer for evaluation.
|
|
It was reported patient had a three lumen picc in right upper arm.During dressing change; lines were checked for patency.White line was found to be leaking where a secureacath was placed over the line.The picc team was called to replace picc line.Once the line was replaced, picc team evaluated the old line and found the secureacath had pinched and cause the line to split lengthwise down the line about an 1/2 inch, causing it to leak.No injuries reported.
|