| Model Number |
M00545190 |
| Medical Device Problem Codes |
Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Imdrf impact code f2301 captures the reportable event of additional device required (net).
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 39 was used in the duodenum and common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat stones performed on (b)(6) 2025.During sphincterotomy, the tech bowed the tome slightly and the cut wire popped off the catheter.The detached cut wire was successfully retrieved using a net.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Imdrf impact code f2301 captures the reportable event of additional device required (net).Block h11: investigation results the returned autotome rx 39 was analyzed, and a visual and microscopic inspection found that cutting wire was blackened, kinked and broken at 8 mm from the proximal pierce hole.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of cutting wire break was confirmed.The results of the analysis performed on the returned device found that the cutting wire was kinked and broken at 8 mm from the proximal pierce hole.Based on the condition of the device, the problem found could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire's integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can bent the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 39 was used in the duodenum and common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat stones performed on (b)(6) 2025.During sphincterotomy, the tech bowed the tome slightly and the cut wire popped off the catheter.The detached cut wire was successfully retrieved using a net.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: imdrf device code a0401 captures the reportable event of cutting wire broken.Imdrf impact code f2301 captures the reportable event of additional device required (net).H11: investigation results: the returned autotome rx 39 was analyzed, and a visual inspection found that cutting wire was blackened, kinked and broken at 8 mm from the proximal pierce hole.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of cutting wire break was confirmed.The results of the analysis performed on the returned device found that the cutting wire was kinked and broken at 8 mm from the proximal pierce hole.Based on the condition of the device, the problem found could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire's integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can bent the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 39 was used in the duodenum and common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat stones performed on (b)(6) 2025.During sphincterotomy, the tech bowed the tome slightly and the cut wire popped off the catheter.The detached cut wire was successfully retrieved using a net.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
|
| |
|