| Model Number |
EEAORVIL25A |
| Medical Device Problem Code |
Incomplete or Inadequate Connection (4037)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant medical products: eea2535, eea2535 eea 25 mm-3.5 sgl use stapler (lot#:p2k0417) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a laparoscopic roux-en-y procedure, the device was being applied to the anastomosis.However the oral anvil would not connect to the stapler.To resolve the issue a new stapler and oral anvil were used.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d3 (mfr name and address), d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the orvil anvil was observed to be tilted and separated from the instrument.It was reported that the oral anvil would not connect to the stapler.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the transoral circular stapler anvil is compatible with the corresponding diameter eea¿ xl stapler length products only.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|