H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.Imaging evaluation summary: the images received cannot be used to perform a full imaging evaluation because they do not meet the dicom standard.The extent and accuracy of the observations and findings may be limited due to the completeness, format and/or quality of the images provided for review.Gore cannot guarantee the images provided are complete, accurate or lack alteration.Therefore, gore cannot guarantee all key findings have been captured or that the findings are accurate.The imaging evaluation performed by a clinical imaging specialist showed the following: three radiographic jpeg images provided for evaluation.No name, date, or demographics on the images.Cannot manipulate the images in any way.(window leveling, rotating, etc.) cannot provide diameter or length measurements with jpeg images.Jpeg #1 shows: there appears to be a partial occlusion on the side of the contra gate.So this involves the ipsilateral trunk device and contralateral leg, proximally.Devices distal to the contra ring appear to be completely occluded.So this includes the distal portion of the contralateral leg, the iliac branch component, and the internal iliac component.Jpeg #2 shows: 3d reconstruction image this image confirms the ibe device along with the internal iliac component is implanted on the patient¿s right side.Jpeg #3 shows: there appears to be an additional stent implanted within the previously implanted contralateral leg, ceb, and into the external iliac artery.There is flow within the added leg on the patient¿s right side and in the devices implanted on the patient¿s left side.The right internal iliac artery and the device there in appears to be covered.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
The following information was reported to gore: on (b)(6) 2021, this patient underwent endovascular treatment for the right common iliac artery aneurysm using gore excluder aaa endoprostheses and gore excluder iliac branch endoprostheses.A total of five stent grafts were implanted, including one trunk ipsilateral leg, two contralateral legs, one iliac branch component and one internal iliac component.In (b)(6) 2025 (exact date unknown), a follow-up ct scan revealed a thrombotic occlusion extending from the bifurcation of the trunk ipsilateral leg to the right groin area.On (b)(6), 2025, a reintervention was performed including a thrombectomy.Subsequently, one stent graft was additionally implanted in each of the left and right common iliac arteries.Another stent graft was implanted in the right external iliac as an extension, and the right internal iliac was intentionally covered.After expanding the pta balloon in the right distal area, a bare metal stent was placed near the right groin.The patient tolerated the procedure.The physician¿s comment: ¿the reason for the thrombotic occlusion is unknown.¿.
|