| Model Number |
EM800 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Repair request initiated for device with the report of overheating.There was no patient involvement.Repair is being escalated to product event due to reason for the return.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis: no conclusion can be drawn, device evaluation anticipated, but not yet begun.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis :evaluation on overheating couldn't able to perform due to attachment and tool both are not locking.It was also found that motor is not locking, found noisy, hipot and depth test failed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Repair request initiated for device with the report of overheating.There was no patient involvement.Repair is being escalated to product event due to reason for the return.
|
| |
|
Search Alerts/Recalls
|