ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF MULTIPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number MA60AC
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-health care professional reported that during an intraocular lens (iol) implant procedure, noticed that the lens was broken, with no patient contact.The procedure was completed on same day without any harm to the patient.
 
Additional Manufacturer Narrative
Additional information was provided in h.3., h.6.And h.11 the product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The account indicated the use of a qualified cartridge and handpiece with a non-qualified viscoelastic.The root cause for the reported complaint may be related to a failure to follow the instruction for use (ifu).The viscoelastic indicated is not qualified for the lens/company combination used.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: use holding forceps to grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ophthalmic viscosurgical device (ovd) filled cartridge.The lens should be inserted until it is centered with or slightly past the outline etched into the top of the cartridge.Lenses with a 6.5 mm optic diameter may need to be pre-folded for easy entry into the back of the cartridge.Lens may be pre-folded by gently applying pressure to the edge of the lens while holding the central optic with forceps the trailing haptic will extend from the proximal end of the cartridge.Position the trailing haptic to the left of the haptic post as shown in the detailed view.This will allow the plunger to go past the haptic during the initial advancement of the plunger into the cartridge.Verify the lens is positioned on the bottom surface of the cartridge.The haptic should be maintained to the left of the post when the lens is loaded correctly.Accurate positioning of the lens will decrease the potential for optic and haptic damage.Due diligence has been performed in an attempt to obtain further information on this event.The account indicated the product was not available to evaluate.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF MULTIPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22403630
Report Number1119421-2025-01840
Device Sequence Number13256366
Product Code HQL
UDI-Device Identifier00380652552523
UDI-Public(01)00380652552523(17)260418(11)210420(10)15187155
Combination Product (Y/N)N
Initial Reporter CountryAR
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/18/2026
Device Model NumberMA60AC
Device Catalogue NumberMA60AC.200
Device Lot Number15187155
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/10/2025
Supplement Date Received by Manufacturer08/01/2025
Initial Report FDA Received Date07/04/2025
Supplement Report FDA Received Date08/26/2025
Date Device Manufactured04/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH II IOL DELIVERY SYSTEM, CARTRIDGE B; MONARCH III IOL DELIVERY SYSTEM, INJECTOR; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE
Patient SexUnknown
-
-