ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SRF-605
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 06/03/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
 
Event or Problem Description
Healthcare professional reported "capsular contracture, baker grade iii".This record is for the right side.Device was explanted and replaced.
 
Event or Problem Description
Healthcare professional reported "capsular contracture, baker grade iii".This record is for the right side.Device was explanted and replaced.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation: based on the product analysis performed, the assessments of the complaints are: capsular contracture: unable to observe as it is not related to the manufacturing process.As per the investigation procedures, creases and deformation were completed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key22403639
Report Number9617229-2025-11129
Device Sequence Number15541939
Product Code FTR
UDI-Device Identifier10888628007383
UDI-Public(01)10888628007383(10)3465956(11)201005(17)250905
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/05/2025
Device Catalogue NumberSRF-605
Device Lot Number3465956
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/10/2025
Supplement Date Received by Manufacturer08/12/2025
Initial Report FDA Received Date07/04/2025
Supplement Report FDA Received Date08/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/05/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexFemale
Patient Weight83 KG
Patient EthnicityHispanic
Patient RaceWhite
-
-