| Model Number |
BF-P150 |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Unstable (1667)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the results of the investigation and previous investigations through the product technical investigation (pti) process, reasonably known information suggests probable causes due to some kind of stress such as damage or deterioration of parts.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the bronchovideoscope exhibited an unsecured distal end due to adhesive peeling off the distal end.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|