EXACTECH, INC. PS TIBIAL INSERTS SZ 2, 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 204-22-13
Medical Device Problem Codes Fracture (1260); Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Osteolysis (2377); Inadequate Osseointegration (2646)
Date of Event 12/20/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10 concomitants: (b)(6) 200-04-22 - (disc) cemented finned t ib.Tra sz 2f/2t.(b)(6) 234-02-02 - (disc) optetrak asy,ps c emented femoral, sz 2.(b)(6) 21-0901-19l-v1 - hall lpi90x13/21x1.19 nueva (v193-119-90).(b)(6) 21-0901-19y-v1 - stryker lpi s7/6/5/4 h.Sierra 90*13/21*1,19.H3: the device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that this patient's left knee was revised approximately 8 years 6 months post initial operation.Patient was experiencing premature wear of the polyethylene with mechanical loosening of the implant secondary to massive osteolysis.Patient was revised to a competitor's device.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: h6, h11.The following sections were corrected: h6.H11: the reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/technique.A review of manufacturing data was performed, and no discrepancies were found.A definitive root cause was unable to be determined; however, was likely due to the reported loosening, prosthesis wear, or due to the inclusion of the polyethylene in the packaging recall.Failure of the cement used to secure the femoral and/or tibial component to the respective bone, may have lead to loosening at the cement-implant and/or cement-bone interface; however, this cannot be confirmed.Possible causes for polyethylene wear include malalignment between the implants, high contact stresses during knee flexion, excessive posterior slope, third body wear, patient-related conditions, instability, or any combination of these possibilities.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS TIBIAL INSERTS SZ 2, 13MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key22411232
Report Number1038671-2025-02426
Device Sequence Number21251017
Product Code JWH
UDI-Device Identifier10885862048240
UDI-Public10885862048240
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/15/2018
Device Catalogue Number204-22-13
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2025
Supplement Date Received by Manufacturer10/16/2025
Initial Report FDA Received Date07/07/2025
Supplement Report FDA Received Date11/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/16/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization;
Patient SexFemale
-
-