EXACTECH, INC. PS TIBIAL INSERTS SZ 2, 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
204-22-13 |
| Medical Device Problem Codes |
Fracture (1260); Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Osteolysis (2377); Inadequate Osseointegration (2646)
|
| Date of Event |
12/20/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10 concomitants: (b)(6) 200-04-22 - (disc) cemented finned t ib.Tra sz 2f/2t.(b)(6) 234-02-02 - (disc) optetrak asy,ps c emented femoral, sz 2.(b)(6) 21-0901-19l-v1 - hall lpi90x13/21x1.19 nueva (v193-119-90).(b)(6) 21-0901-19y-v1 - stryker lpi s7/6/5/4 h.Sierra 90*13/21*1,19.H3: the device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that this patient's left knee was revised approximately 8 years 6 months post initial operation.Patient was experiencing premature wear of the polyethylene with mechanical loosening of the implant secondary to massive osteolysis.Patient was revised to a competitor's device.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: h6, h11.The following sections were corrected: h6.H11: the reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/technique.A review of manufacturing data was performed, and no discrepancies were found.A definitive root cause was unable to be determined; however, was likely due to the reported loosening, prosthesis wear, or due to the inclusion of the polyethylene in the packaging recall.Failure of the cement used to secure the femoral and/or tibial component to the respective bone, may have lead to loosening at the cement-implant and/or cement-bone interface; however, this cannot be confirmed.Possible causes for polyethylene wear include malalignment between the implants, high contact stresses during knee flexion, excessive posterior slope, third body wear, patient-related conditions, instability, or any combination of these possibilities.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|