| Catalog Number |
367341 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D.2b medical device type: one additional code applies; jka.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd vacutainer® push button blood collection set, the rubber covering the non-patient needle had a bulge causing blood to leak from the needle while drawing blood on unspecified number of devices.There was no health impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd had not received any samples for investigation.Therefore, functional tests were conducted on 30 retained samples using 10 tubes per needle to evaluate the device's functionality.All samples passed the tests with no evidence of side pierced sleeves, poor sleeve recovery, or sleeve leakage during the draw.Additionally, these 30 retained samples underwent another set of functional tests focused on retraction lockout and all samples passed these tests as well, demonstrating proper activation with no evidence of defects or leakage.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: sleeve leakage.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd vacutainer® push button blood collection set, the rubber covering the non-patient needle had a bulge causing blood to leak from the needle while drawing blood on unspecified number of devices.There was no health impact or consequence reported.
|
| |
|