• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE G; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE G; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 0185
Medical Device Problem Codes Signal Artifact/Noise (1036); Failure to Capture (1081)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system was exhibiting intermittent right ventricular (rv) lead loss of capture.Technical services (ts) was consulted and noted electromagnetic interference (emi), the patient has a left ventricular assist device.This crt-d system remains in service.No adverse patient effects were reported.Additional information was received and there has not been right atrial (ra) lead capture for several months, and the right ventricular (rv) lead was reported to not be working well.Technical services (ts) was consulted and recommended reprogramming options.Tachy therapy had been turned off.This crt-d system remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE G
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
#12, rd. #698
dorado PR 00646
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key22413123
Report Number2124215-2025-44828
Device Sequence Number8293661
Product Code LWS
UDI-Device Identifier00802526531194
UDI-Public00802526531194
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P910073/S043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2004
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/18/2006
Device Model Number0185
Device Catalogue Number0185
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2025
Initial Report FDA Received Date07/07/2025
Date Device Manufactured02/18/2004
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age72 YR
Patient SexMale
-
-