Return qty.1, 24050, 30k fsi-sli-10s fg-ko, 00114639 bul per manufacture date.Test method / gp005-wi02.Inductance: gauge id#: 5349, due: 12/31/2026, result 3.09, meets spec, does not meet spec, n/a, tip condition: meets spec, does not meet spec, n/a, stack condition, meets spec, does not meet spec, n/a, tested on: g130 gage id#: 9626-2, due date: 09/30/2025, set pressure: 35 psi, water flow: output rate: 35 psi, meets spec, does not meet spec, n/a, spray: meets spec, does not meet spec, n/a, water leak: hp sealing ring, proximal ring, distal ring, seam leak, n/a, vibration: meets spec, does not meet spec, n/a, grip condition: meet, does not meet spec, n/a.Other visual observation: insert tip temperature measured and is within acceptable range.It was observed that the grip is loose but not rotating around the connecting body.Insert was tested on a digital thermometer i.D.#: 6116a.Due date: 09/30/2025.The temperature was 98.9 °f, no fault found with temperature.The specifications state the temperature is not to exceed 118.4°f.(per pds-8016 rev 6).Alleged event: confirmed, not confirmed.
|