MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL CERENOVUS ENTERPRISE; INTRACRANIAL NEUROVASCULAR STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL CERENOVUS ENTERPRISE; INTRACRANIAL NEUROVASCULAR STENT Back to Search Results
Catalog Number ENCR402312
Medical Device Problem Codes Premature Activation (1484); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer ref#: (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section e1.Initial reporter phone: (b)(6).Section h3 - the device is available to be returned for evaluation and testing.However, it has not been received to date.If the device returns, a device investigation will be performed.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
A healthcare professional reported that during an endovascular embolization, an enterprise2 4mmx23mm vascular reconstruction device (encr402312, 9104195) was impeded in the y connector and could not advance any more.The doctor only retracted the stent alone, but the stent was found detached from the delivery wire in the y connector.The doctor removed they connector and removed the stent and then switched to a new stent to complete the surgery.The microcatheter was not replaced.There was no patient injury reported.Additional event information received on 07-jul-2025 indicated that there was no evidence of physical material within the device.The microcatheter used was a prowler select plus 150/5cm microcatheter (606s255x, unknown lot).The microcatheter did not kink/bent.There was a procedure prolongation/delay due to the event, however, the delay did not result in a patient adverse consequence.
 
Additional Manufacturer Narrative
Manufacturer ref# (b)(4).Updated sections on this medwatch: b4, d9, g3, g6, h2, h3, h6 and h11.Complaint conclusion: a healthcare professional reported that during an endovascular embolization, an enterprise2 4mmx23mm vascular reconstruction device (b)(6) was impeded in the y connector and could not advance any more.The doctor only retracted the stent alone, but the stent was found detached from the delivery wire in the y connector.The doctor removed the y connector and removed the stent and then switched to a new stent to complete the surgery.The microcatheter was not replaced.There was no patient injury reported.Additional event information received on 07-jul-2025 indicated that there was no evidence of physical material within the device.The microcatheter used was a prowler select plus 150/5cm microcatheter (b)(6).The microcatheter did not kink/bent.There was a procedure prolongation/delay due to the event, however, the delay did not result in a patient adverse consequence.The device was returned to j&j medtech for further evaluation.A non-sterile enterprise2 4mmx23mm vascular reconstruction device was received contained in the decontamination pouch.Upon receiving the device, visual inspection was performed, and as reported, the stent component was already detached from the delivery system.The delivery wire was found damaged at the distal end of the proximal coil.No other damages were noted.Microscopic inspection was performed on the stent component.It was observed to be in good condition; there was no structural damage (i.E., no broken struts, no kinks); also, it was noted fully expanded, both ends can be noted as completely flared.The delivery wire was subjected to dimensional analysis and all measurements were found to be within specification, including those specifications that control the attachment and delivery of the stent.Therefore, device failure is not suspected to be a contributing factor.A device history record (dhr) was performed and it indicated this product was final inspection tested at lake region medical and was determined to be acceptable.Although the stent detachment prevented proper analysis for the impeded condition reported, the customer complaint is confirmed based on the findings mentioned above.It is suggested that excessive force could had been inadvertently applied during the attempts to advance the stent through the y-connector, resulting in the damaged condition of the delivery wire and detached condition of the stent.With the limited information available, a conclusive cause cannot be determined; however, it is possible that clinical and procedural factors, including device manipulation and operator's technique, may have contributed to the reported failure.At this time, there is no evidence to support that the issue reported in the complaint is a result of a defect inherently related to the device.As part of the j&j medtech quality process, all devices are manufactured, inspected, and released to approved specifications.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no capa activity is required.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) do contain the following recommendations do not partially deploy the stent from the introducer.Do not apply undue force if resistance is encountered at any point during stent manipulation.Withdraw the unit and advance a new one.Confirm the tip of the delivery wire is entirely within the introducer.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERENOVUS ENTERPRISE
Common Device Name
INTRACRANIAL NEUROVASCULAR STENT
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC
325 paramount dr
raynham MA 02767
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
9497882078
MDR Report Key22415420
Report Number3008114965-2025-00683
Device Sequence Number8293709
Product Code NJE
UDI-Device Identifier10886704075356
UDI-Public(01)10886704075356(17)270528(10)9104195
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
H60001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberENCR402312
Device Lot Number9104195
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2025
Supplement Date Received by Manufacturer07/29/2025
Initial Report FDA Received Date07/07/2025
Supplement Report FDA Received Date08/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PROWLER SELECT PLUS 150/5CM
Patient SexUnknown
-
-