|
An outside united states (ous) customer contacted a siemens customer care center to report the observation of imprecise hcg results obtained on diluted patient samples on an immulite 2000 xpi instrument.Quality controls (qcs) recovered within ranges on the day of the event.A siemens customer service engineer (cse) was dispatched to the customer¿s site.During this visit, the cse checked the system.The cse adjusted the sample and reagent probes, the drd, and the dilution well motor speed.The pmt shutter was checked.A water test was performed and the results were acceptable.Per the limitations section in the immulite 2000 hcg instructions for use (ifu): "for diagnostic purposes, the results obtained from this assay should always be used in combination with the clinical examination, patient medical history, and other findings." siemens is investigating.
|
|
Siemens filed the initial mdr 1219913-2025-00137 on 07-jul-2025.Additional information 28-oct-2025: a siemens customer service engineer (cse) was dispatched to the customer¿s site.During this visit, the cse checked the system.The cse adjusted the sample and reagent probes, the drd, and the dilution well motor speed.The pmt shutter was checked.A water test was performed and the results were acceptable.Siemens evaluated the information / data provided.The customer data included two adjustments during the relevant time period.It was noted that the adjustments had an increased intercept.However, the slope was within range and the adjustments were complete.Biorad quality control (qc) level 1 and level 2 were within range, near the mean value.To further evaluate the reported complaint the customer was asked how the dilutions were performed, onboard or manually diluted.The customer reported that the dilutions performed are onboard dilutions.For further troubleshooting, the customer was asked to perform onboard dilutions compared to a corresponding manual dilution with the samples affected.For example, 1:10 and 1:20 onboard dilution compared to manual dilution 1:10 and 1:20.The customer provided the following data: auto diluted 1:10 - auto diluted 1:20 - manual diluted 1:10 - manual diluted 1:20 patient 1, 14477, 14566, 15571, 17296.Patient 2, >50000, 73301, >50000, 75415.Patient 3, 16615, 19811, 18970, 19241.Patient 4, >50000, 55457, >50000, 49209.Patient 5, 12213, 16994, 12024, 13474.No clear pattern is visible for the different dilution results provided.The precision is within specifications.A service visit was recommended to verify the dilution well, as well as the sample handling.However, the customer opted not to proceed with a service visit at this time.Currently, the customer is not reporting any concerns regarding the results and, therefore, do not wish to initiate any intervention on the instrument.The customer has been advised to file a new complaint if this issue reoccurs.No additional complaints have been received for immulite 2000 hcg kit lot 483 and imprecision on diluted samples.The customer is operational.Based on the available information, the probable cause of the discordant results is consistent with sample integrity issues.The immulite 2000 hcg kit lot 483 is performing as intended.A product performance issue has not been identified.No further evaluation of the device is required.In section h6, the investigation findings and investigation conclusions codes were updated.
|