| Model Number |
406968 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the atrial fibrillation procedure, the sheath was inserted into the heart.After the viewflex catheter was inserted and then removed, blood leaked from the hemostasis valve when no catheter was inserted.The sheath was replaced, and the procedure was completed with no adverse consequences to the patient.No air movement into the patient was confirmed.The only products inserted directly into the hemostasis valve were the included dilator and viewflex catheter.No additional venipuncture was performed due to sheath replacement.There was no resistance when inserting the dilator.The sheath and dilator were integrated and inserted as per the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
One 10f fast-cath introducer sheath was received for evaluation.Functional testing determined a leak was noted in the hemostasis valve.The cap was removed from the hemostasis hub and the hemostasis seal was microscopically inspected.Tearing was noted on both sides of the seal.A corrective action was initiated to address the failure mode of this introducer leak.
|
| |
|