Catalog Number TSVH11
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Pain (1994)
Date of Event 06/02/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant at tooth site #4 fractured and was removed.Patient presented for an evaluation of implant #4 and it was determined that the implant had fractured, so the patient was scheduled to have the implant removed.Symptoms as a result of the event: pain.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: patient weight unknown / not provided.G4: additional pma/510(k) number k013227.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
Zimvie received one (1) tsvh11, (impl tapered scr-v ha 3.7 mm 3.5mm 11.5mm) for evaluation.Visual evaluation was performed, the implant has signs of use.The implant had bone debris on external threads.The was a hairline fracture identified on the implant's collar.Dhr review was completed for the subject lot number 1248168.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1248168 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿dental: functional: fracture: implant¿.The customer did not submit images for the reported event.Based on the investigation and risk management file review as per rm-00541-haz, the most likely root cause determined from the investigation was patient factors.Therefore, based on the available information, a device malfunction did occur.The implant was fractured at the collar.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report.D9: device availability.G3: date received by manufacturer.G6: checked "follow-up".H2: checked follow-up type.H3: changed "no" to "yes".H6: entered evaluation codes.H10: added manufacturer narrative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key22420893
Report Number0002023141-2025-01827
Device Sequence Number18347475
Product Code DZE
UDI-Device Identifier00889024019577
UDI-Public(01)00889024019577(17)261112(10)1248168
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K061410
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Catalogue NumberTSVH11
Device Lot Number1248168
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/07/2025
Supplement Report FDA Received Date09/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/13/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-