Correction: h3.Yes.H6.Investigation findings: 3252.Investigation conclusions: 61.Device evaluation: visual inspection confirms that the distal hexalobe of the driver has sheared off.The root cause is likely excessive force applied to the driver tip during insertion of advancement of a screw.Review of device history records showed no manufacturing or processing related irregularities that would cause or contribute to this event.Labeling review: warnings/cautions/precautions: risks identified with the use of these devices, which may require additional surgery, include device component failure, loss of fixation/stabilization, non-union, vertebral fracture, neurological injury, vascular or visceral injury.Possible adverse effects: initial or delayed loosening, disassembly, bending, dislocation, and/or breakage of device components.
|