| Model Number |
5054-52 |
| Medical Device Problem Codes |
Pacing Problem (1439); Extracardiac stimulation (1463); Impedance Problem (2950); Adverse Event Without Identified Device or Use Problem (2993); High Capture Threshold (3266)
|
| Health Effect - Clinical Codes |
Device Overstimulation of Tissue (1991); Discomfort (2330); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient experienced discomfort and pocket stimulation.It was reported that the right ventricular (rv) lead exhibited varying impedances, a large drop in impedance and high thresholds.An increase in rv pacing and high outputs was also observed.The rv lead and implantable pulse generator (ipg) remain in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 5076-45 lead implanted on (b)(6) 2004.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance trend of the right ventricular pacing lead was variable.Analysis of the device memory indicated the impedance trend of the right ventricular pacing lead was decreasing.Analysis of the device memory indicated pacing capture threshold in the right ventricle was elevated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was further reported that there was no performance issue with the ipg.The rv lead outputs were reprogrammed.It was also noted that the impedance had stabilized.
|
| |
|