H6: a medtronic representative went to the site to test the equipment.Testing revealed that replaced collimator and tested, ready for use.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: bi71000874, medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Medtronic received information regarding an imaging system being used outside of the procedure.It was reported that when booting up the system, the pendant was displaying the collimation icon on the screen.Clinical specialist (cs) went to open the collimation but it wouldn't open so cs rebooted system.Upon boot-up, cs was receiving an error "error initializing collimator." there was no patient involved.
|