Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional manufacturer narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
|
|
It was reported that the sterilization department of the hospital noticed, during inspection of the instruments, some stains on several the instruments for augmentation (distal & posterior) of the femoral component.It looks like there is some rust formation on the edge.There were no patient consequences.
|
|
Product complaint # (b)(4).Investigation summary = > according to the information received, "the sterilization department of the hospital saw during inspection of the instruments some stains on several the instruments for augmentation (distal & posterior) of the femoral component.It looks like there is some rust formation on the edge of the magnet (in the augment)" the product was not returned to medtech orthopaedics; however, photos were provided for review.See attachment dpnl25-048 picture11,dpnl25-048 picture10,dpnl25-048 picture9,dpnl25-048 picture8,dpnl25-048 picture7,dpnl25-048 picture6,dpnl25-048 picture5,dpnl25-048 picture4, dpnl25-048 picture3,dpnl25-048 picture2,dpnl25-048 picture1,dpnl25-048 picture20,dpnl25-048 picture19,dpnl25-048 picture18,dpnl25-048 picture17,dpnl25-048 picture16,dpnl25-048 picture15,dpnl25-048 picture14,dpnl25-048 picture13 & dpnl25-048 picture12.The photo investigation revealed that attun rev dst f augtrl 8xsz3-4 had foreign substances.The observed condition is likely due to improper cleaning/sterilization.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the attun rev dst f augtrl 8xsz3-4 would contribute to the complained device issue.Based on the investigation findings, potential cause attributed to improper maintenance and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the device lot number is unknown, therefore a¿device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.
|