MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 8300AB
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/12/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: additional manufacturer narrative: the subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was reported that a patient with a 19mm 8300ab aortic valve underwent a valve-in-valve procedure after an implant duration of 5 years, 3 months due to unknown reason.The tavr was successfully performed with a 20mm 9750tfx valve without issues and the patient is stable.
 
Additional Manufacturer Narrative
H10: additional manufacturer narrative: updated sections: d4 (expiration date), g3, g6, h2, h4, h6 (investigation findings, and investigation conclusions).Due diligence attempts were made to gather additional information regarding a device failure mode, however additional information is not available at this time.Patient medical records were not provided by the hospital for review, or the medical records provided do not clarify the device failure mode.In this case, there is no evidence of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no use-related issue with a hazardous situation; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.Per the instructions for use (ifu), reoperation/explant are known potential adverse events associated with bioprosthetic cardiac valves and annuloplasty rings.Based on the information available, a definitive root cause cannot be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE VALVE SYSTEM
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key22437634
Report Number2015691-2025-05532
Device Sequence Number17310816
Product Code LWR
UDI-Device Identifier00690103194470
UDI-Public(01)00690103194470(17)220507
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/07/2022
Device Model Number8300AB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2025
Supplement Date Received by Manufacturer07/29/2025
Initial Report FDA Received Date07/08/2025
Supplement Report FDA Received Date07/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/08/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Life Threatening; Hospitalization;
Patient Age71 YR
Patient SexFemale
-
-