• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 LUBRI-SIL® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 LUBRI-SIL® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY Back to Search Results
Catalog Number 919214A
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/26/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/eporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during pre-test the foley catheter sterile water did not drain.
 
Event or Problem Description
It was reported that during pre-test the foley catheter sterile water did not drain.Per notification from ucc on 17dec2025, it was reported that asymmetrical balloon was found during sample evaluation on 05nov2025.
 
Additional Manufacturer Narrative
The reported event is confirmed via capa associated.This is addressed by capa 12474540.Photo samples were submitted showing the catheter balloon inflated asymmetrically.Unable to confirm whether the catheter was fully inflated from the photos.The silicone foley catheter was returned with the syringe.Visual evaluation noted the catheter balloon was inflated and asymmetrical.The catheter balloon was inflated with 10ml methylene blue solution (mbs) using the returned syringe.The balloon inflated with no difficulty and was noted to inflated asymmetrically (100:0); this failure is documented and investigated via capa 11291030.An attempt was made to deflate the catheter balloon, however, only 5ml deflated in 5 minutes.The catheter balloon was then inflated with 10ml mbs using an in-house syringe, the balloon passively deflated in 01min52sec.The reported event is confirmed against the returned syringe.Capa 12474540 identified the root cause of this failure as a combination of three (3) factors; provided water syringe does not meet user expectation for smooth engagement with the valve and no instructions are provided in the japanese ifu for aspirating to assist in deflation, deflation time or take off force criteria or specification to deflate the catheter has not been determined and it is unknown for the customers, inadequate process qualification performed with change in syringe supplier and mold change.Risk review, labeling review, and dhr review are not required as event has already been investigated per capa 12474540, however, dhr review was completed prior to capa determination.Correction: d, f, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUBRI-SIL® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY
Common Device Name
TEMPERATURE-SENSING URINE METER FOLEY TRAY
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key22438030
Report Number1018233-2025-05644
Device Sequence Number11545046
Product Code MJC
UDI-Device Identifier00801741040122
UDI-Public(01)00801741040122(11)250220(17)270628(10)MYJZ4529
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number919214A
Device Lot NumberMYJZ4529
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2025
Supplement Date Received by Manufacturer12/18/2025
Initial Report FDA Received Date07/08/2025
Supplement Report FDA Received Date12/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-