| Catalog Number |
ES2824 |
| Medical Device Problem Codes |
Inflation Problem (1310); Failure to Deflate (4060)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed in veeva to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information, during the initial implant procedure, when the left side cylinder was placed it would not inflate, and when it was removed it would not deflate.A new cylinder set with pump was implanted to complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan touch pump and cylinders 1 and 2 were received for evaluation.No functional abnormalities were noted with the pump.A separation was noted on the bladder of cylinder 1.This is a site of leakage.The separation has a central groove, indicating contact with sharp instrumentation such as a needle.No functional abnormalities were noted with cylinder 2.Because these components were released according to manufacturing and quality control procedures, it was concluded that the observed instrument separation in the bladder of cylinder 1 occurred subsequent to the device packaging being opened.The information received indicated the issue was noticed during the implant of the device.Based on the evaluation and information received, the separation most likely occurred inadvertently during the implant procedure.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
|
| |
|
Event or Problem Description
|
|
According to the available information, during the initial implant procedure, when the left side cylinder was placed it would not inflate, and when it was removed it would not deflate.A new cylinder set with pump was implanted to complete the procedure.
|
| |
|