• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FERNO-WASHINGTON, INC. INX FOR INLINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FERNO-WASHINGTON, INC. INX FOR INLINE Back to Search Results
Model Number 0015811
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 05/25/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported while loading the stretcher into the ambulance via manual mode, due to a dead battery, one of the medic's got their hand caught between the stretcher frame and legs resulting in a soft tissue injury.The medic sought medical treatment and required time off work; however, followup details of the injury were not provided.This incident occurred in canada.There was no report of injury or adverse effect on the patient.The end user further advised their internal investigation resulted in a finding of unintended use error.An authorized service technician did evaluate the stretcher and battery and both were found to be functioning as intended and no repairs were required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INX FOR INLINE
Common Device Name
INX FOR INLINE
Manufacturer (Section D)
FERNO-WASHINGTON, INC.
70 wiel way
wilmington OH 45177
Manufacturer (Section G)
FERNO-WASHINGTON, INC.
70 wiel way
wilmington OH 45177
Manufacturer Contact
dawn greene
70 wiel way
wilmington, OH 45177
9372832900
MDR Report Key22438957
Report Number1523574-2025-00066
Device Sequence Number14241098
Product Code FPO
UDI-Device Identifier00190790001339
UDI-Public(01)00190790001339(11)200608
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number0015811
Device Catalogue Number0015811
Device Lot NumberM707100
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2025
Initial Report FDA Received Date07/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/08/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-