As reported by an edwards lifesciences field clinical specialist, approximately 10 years and 5 months post implantation of a sapien xt valve in the aortic position, a valve in valve with a 26mm sapien 3 ultra resilia valve was performed successfully due to central regurgitation.
|
|
A supplemental mdr is being submitted due to engineering evaluation findings.Sections: g3, g6, h2, h6 type of investigation, investigation findings, investigation conclusions have been updated.The event reported is anticipated in the risk management documentation for transcatheter heart valve procedures.Additional assessment of the failure mode is not required at this time.The device was not returned for evaluation.Due to the unavailability of the complaint device, engineering was unable to perform any visual inspection, functional testing, or dimensional analysis.The instructions for use (ifu), and training manuals have been reviewed.No inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported central leak was confirmed based on reported of fcs (field clinical specialist).The reported event does not allege a malfunction that could be related to an edwards manufacturing deficiency.A review of the ifu revealed no deficiencies.During the manufacturing process, all sapien xt valves are 100% visually inspected for defects and 100% tested for proper coaptation under physiological backpressure conditions prior to release for distribution.Therefore, it is highly unlikely that a manufacturing defect or device malfunction contributed to the event.Per the instructions for use (ifu), valve regurgitation is potential adverse event associated with bioprosthetic heart valves and the tavr procedure.Regurgitation which develops progressively over time can be due to a number of issues including patient related factors or structural valve deterioration, including calcification, non-calcific degeneration, leaflet thickening or fibrosis, or a combination of these.Regurgitation may also develop progressively if host fibrotic tissue, or pannus, grows onto the bioprosthetic valve.Pannus, a cause of nonstructural dysfunction, may interfere with functionality of the device by restricting the leaflet motion leading to abnormal coaptation.Due to limited information, a definitive root cause was unable to be determined at this time.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.
|