• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN? X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN? X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR Back to Search Results
Model Number G146
Medical Device Problem Code Failure to Capture (1081)
Health Effect - Clinical Codes Device Overstimulation of Tissue (1991); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the health care professional (hcp) requested review of the presenting electrogram stored by the cardiac resynchronization therapy defibrillator (crt-d) system.Technical services (ts) reviewed per request.Ts advised there is intermittent loss of capture on the left ventricular channel.The hcp provided the patient would be seen in the clinic for an evaluation.Additional information has been requested but was not provided at this time.This crt-d remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the health care professional (hcp) requested review of the presenting electrogram stored by the cardiac resynchronization therapy defibrillator (crt-d) system.Technical services (ts) reviewed per request.Ts advised there is intermittent loss of capture on the left ventricular (lv) channel.The hcp provided the patient would be seen in the clinic for an evaluation.Additional information from the field representative provided the patient was subsequently seen in clinic.Reprogramming was completed.During reprogramming in some configurations the patient experienced phrenic nerve stimulation.This crt-d remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN? X4 CRT-D
Common Device Name
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112 5798
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key22440540
Report Number2124215-2025-45446
Device Sequence Number11939194
Product Code NIK
UDI-Device Identifier00802526534577
UDI-Public(01)00802526534577(17)170826
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/26/2017
Device Model NumberG146
Device Catalogue NumberG146
Device Lot Number371168
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2025
Supplement Date Received by Manufacturer08/04/2025
Initial Report FDA Received Date07/08/2025
Supplement Report FDA Received Date08/12/2025
Date Device Manufactured08/26/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age80 YR
Patient SexMale
-
-