ZIMMER DENTAL IMPL TAPERED SCR-V SBM 6M M 5.7MM 10MM; DENTAL IMPLANT Back to Search Results
Catalog Number TSV6B10
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Unspecified Infection (1930); Pain (1994)
Date of Event 06/12/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented last week (not yet restored) with pain, infection and mobility of the implant at tooth site #30.The implant was removed due to allergic reaction and infection.Symptoms as a result of the event: abscess.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A1: patient identifier unknown / not provided.A2: age and date of birth unknown / not provided.A4: weight unknown / not provided.G4: additional pma/510(k) number ¿ k011028 | k013227.Product has been received by zimvie and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 67.H10: narrative/data was updated.Zimvie received the reported implant for evaluation.Visual evaluation was performed, the implant had signs of use with debris on its internal and external threads.The implant has markings from removal.There was no damage identified or signs of malfunction that would have contributed to the event.Measurements match drawing.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was patient factors.A summary investigation has been completed for infection events identifying that a definitive root cause cannot be identified due to the wide range of external (non-design/ non-manufacturing related) factors potentially impacting the sterility of the implant and its environment.Therefore, based on the available information, device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL TAPERED SCR-V SBM 6M M 5.7MM 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key22441287
Report Number0002023141-2025-01851
Device Sequence Number7904529
Product Code DZE
UDI-Device Identifier00889024019379
UDI-Public(01)00889024019379(17)250809(10)1238646
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K061410
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 09/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Expiration Date08/09/2025
Device Catalogue NumberTSV6B10
Device Lot Number1238646
Was Device Available for Evaluation? No
Date Returned to Manufacturer06/20/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/08/2025
Supplement Report FDA Received Date09/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/10/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-