Model Number 33510FS
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/20/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.The device was not yet returned for investigation.Cross reference complaint id: (b)(4).The event is filed under internal karl storz complaint id: (b)(4).
 
Event or Problem Description
It was reported that both forceps inserts broke off before the trocar was inserted.No negative impact in state of health reported.Cross reference mdr: 9610617-2025-00169.
 
Additional Manufacturer Narrative
Result of investigation: the cause of the complaint is not attributable to a production error, as the affected items originate from different production years and there are no comparable complaints.The damage identified (breaks, missing pins, deformations) and the external labeling indicate improper handling and unauthorized repairs by the customer.Responsibility: user/customer.The product was returned for evaluation on 7-16-2025, however this information was accidentally omitted on this report but was reported in the related case (b)(4) on time.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLICKLINE FORCEPS INSERT
Common Device Name
CLICKLINE FORCEPS INSERT
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key22448931
Report Number9610617-2025-01170
Device Sequence Number8578537
Product Code GCJ
UDI-Device Identifier04048551151068
UDI-Public04048551151068
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K935071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number33510FS
Device Catalogue Number33510FS
Device Lot NumberNL05
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2025
Supplement Date Received by Manufacturer10/02/2025
Initial Report FDA Received Date07/08/2025
Supplement Report FDA Received Date10/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-