| Model Number |
SIGMRET |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Tissue Breakdown (2681)
|
| Date of Event |
06/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: sig power sigadaptstnd linear adapter (serial#: (b)(6), sigpshell, sig power sigpshell control shell (lot#:n5b1665y), sigphandle, sig power sigphandle handle (serial#: (b)(6), sigmret, sig power sigmret manual retract tool (lot#:unknown), egia60amt egia 60 artic med thick sulu (lot#:p5c1345) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic procedure, the handle with 243 lives and more than 40 lives on the adapter, fired without issue using the first reload.The second reload cycled without any problems, but jammed near the end of the firing process.An unknown message with a yellow background appeared on the screen, accompanied by a blinking jaw icon.Additionally, the adapter swim lane was black with a green outline.The surgeon restarted the device, disconnected and reconnected the adapter.The handle was also replaced.Two retraction tools were used, both of which broke off during the attempt to release the device.Finally, the trapped piece of stomach had to be dissected from the reload to resolve the situation.The surgeon reinforced the gastric sleeve with laparoscopic sutures.A second adapter was used with the same handle to complete the case.After the procedure, using the first handle, the representative was able to remove the closed reload from the adapter without pressing the blue button (which was slightly retracted).The reload was tested on the second adapter but failed to establish a connection or be recognized.When connecting the first ad apter with the closed reload, the screen displayed the message ¿adapter error¿ with a yellow swim lane accompanied by the number ¿1¿ at the bottom of the screen.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h6.Correction: h3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic procedure, the handle with 243 lives and more than 40 lives on the adapter, fired without issue using the first reload.The second reload cycled without any problems, but jammed near the end of the firing process.An unknown message with a yellow background appeared on the screen, accompanied by a blinking jaw icon.Additionally, the adapter swim lane was black with a green outline.The surgeon restarted the device, disconnected and reconnected the adapter.The handle was also replaced.Two retraction tools were used, both of which broke off during the attempt to release the device.No component fell on the patient's cavity.Finally, the trapped piece of stomach had to be dissected from the reload to resolve the situation.The reload could not be unloaded onto the adapter either.The surgeon reinforced the gastric sleeve with laparoscopic sutures.A second adapter was used with the same handle to complete the case.After the procedure, using the first handle, the representative was able to r emove the closed reload from the adapter without pressing the blue button (which was slightly retracted).The reload was tested on the second adapter but failed to establish a connection or be recognized.When connecting the first adapter with the closed reload, t he screen displayed the message "adapter error" with a yellow swim lane accompanied by the number "1" at the bottom of the screen.It was noted that the procedure was extended by more than one hour.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted the distal end of the adapter was inserted into the patient.The manual retract tool had disengaged lever from the trilobe shaft.It was reported that the retraction tool broke off during the attempt to release the device.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|