• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS MAQUET POWERLED II; LAMP, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS MAQUET POWERLED II; LAMP, SURGICAL Back to Search Results
Model Number ARDPWT209005A
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Getinge became aware of an issue with one of surgical lights - powerled ii.It was stated and confirmed by photographic evidence the paint was peeling from headlight.We decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.
 
Additional Manufacturer Narrative
(b)(6).Additional information will be provided following the conclusion of the investigation.
 
Additional Manufacturer Narrative
Getinge became aware of an issue with one of surgical lights - powerled ii.It was stated and confirmed by photographic evidence the paint was peeling from headlight.We decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.Based on the information collected, it was established that when the event occurred, the surgical lights did not meet its specification, since paint peeling could be considered as technical deficiencies, and in this way the device contributed to event.It is unknown if claimed devices were or were not being used for patient treatment or diagnosis when the event took place.According to subject matter expert at manufacturer¿s, the most probable root cause of this paint degradation would be an incorrect cleaning procedure/non-compatible cleaning product.All maquet sas products comply with: - iec 60601-1ed.2.0 & ed.3.0 general requirements for basic safety and essential performance.- iec 60601-1-41 particular requirements for the safety of surgical luminaires and luminaire for diagnosis.- paint definition pfc066.This procedure defines maquet sas¿s requirements for all painted parts.- disinfection product test: the aim of these tests is to detect any incompatibility with disinfectant.This event is clearly due to the cleaning product.Nevertheless, the deterioration of the paint is far advanced and it must have started years ago.Such troubles should have been detected first by the user during daily and monthly checks.The current cleaning product must be improved as described in the user manual.The damaged parts must be replaced as soon as possible.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
 
Event or Problem Description
Manufacturer's reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET POWERLED II
Common Device Name
LAMP, SURGICAL
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key22455819
Report Number9710055-2025-0000177
Device Sequence Number18221508
Product Code FTD
UDI-Device Identifier03700712407179
UDI-Public(01)03700712407179
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARDPWT209005A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2025
Supplement Date Received by Manufacturer08/01/2025
Initial Report FDA Received Date07/09/2025
Supplement Report FDA Received Date08/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2021
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-