| Model Number |
ACL TIGHTROPE |
| Medical Device Problem Codes |
Break (1069); Use of Device Problem (1670); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament surgery the device ar-1588t (lot: 15378964) blocked when pulling in and shortening.Tension cords ultimately tore during further retraction.The tightrope implant was completely removed from the knee joint and cut off from the graft.The surgeon wanted to use a btb tightrope, but the device could not be threaded during use as per instructions.The device ar-1588btb (lot: 14200696) could therefore not be used.The graft was then fixed femorally with a resorbable screw.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a different device.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening/tensioning.
|
| |
|