MAUDE Adverse Event Report: ARTHREX, INC. ACL TIGHTROPE; SMOOTH FIXATION PIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. ACL TIGHTROPE; SMOOTH FIXATION PIN Back to Search Results
Model Number ACL TIGHTROPE
Medical Device Problem Codes Break (1069); Use of Device Problem (1670); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
It was reported that during an anterior cruciate ligament surgery the device ar-1588t (lot: 15378964) blocked when pulling in and shortening.Tension cords ultimately tore during further retraction.The tightrope implant was completely removed from the knee joint and cut off from the graft.The surgeon wanted to use a btb tightrope, but the device could not be threaded during use as per instructions.The device ar-1588btb (lot: 14200696) could therefore not be used.The graft was then fixed femorally with a resorbable screw.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a different device.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening/tensioning.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACL TIGHTROPE
Common Device Name
SMOOTH FIXATION PIN
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22455995
Report Number1220246-2025-02803
Device Sequence Number15492797
Product Code MBI
UDI-Device Identifier00888867019614
UDI-Public(01)00888867019614(11)250127(17)291231(10)15378964
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K241235
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 11/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberACL TIGHTROPE
Device Catalogue NumberAR-1588T
Device Lot Number15378964
Was Device Available for Evaluation? No
Distributor Facility Aware Date06/11/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/11/2025
Supplement Date Received by Manufacturer11/12/2025
Initial Report FDA Received Date07/09/2025
Supplement Report FDA Received Date11/18/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-