Model Number 8229706
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2021
Type of Reportable Event Malfunction
Event or Problem Description
A distributor reported via manufacturer representative that the tube electric resistance had no function pre-operatively.There was no patient involvement.On follow up the device did not function during set-up.The monitor showed incorrect.Troubleshooting was to change a new product.
 
Additional Manufacturer Narrative
H3: analysis of the device found electrically, the resistance of the tube electrodes from end to end shall be less than 200 ohms, and the actual measurements were as follows: 34.8 ohms (blue with white stripe), 40.2 (red with white stripe), and 32.3 ohms (red), in which all were within specification.However, the solid blue electrode had no reading, which would have resulted in the reported event.Under magnification, the defective electrode had cracks and scratches in the silver ink trace underneath the adhesive coating.A review of the global complaint data showed four other complaints about this lot number.In the returned condition, there was an out-of-specification condition likely related to the complaint.This report is being submitted as part of remediation activities associated with capa # 643143, which is addressing mdr reporting gu idance from fdas 1995 preamble, which ensures complaints related to corrections and recalls which were previously reported to fda under 21 cfr 806 are now reported as mdrs; this is not a new malfunction/event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM® EMG - ENDOTRACHEAL TUBE - TRIVANTAGE®
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
7635265677
MDR Report Key22456716
Report Number1045254-2025-01698
Device Sequence Number18197259
Product Code ETN
UDI-Device Identifier00643169782587
UDI-Public(01)00643169782587(17)240406(10)0219861936
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K112686
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification,Recall
Type of Report Initial
Report Date (Section B) 07/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/06/2024
Device Model Number8229706
Device Catalogue Number8229706
Device Lot Number0219861936
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2021
Initial Report FDA Received Date07/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0743-2022
Patient Sequence Number1
Patient SexUnknown
-
-