• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDOPATH*5MM CURVED SCISSORS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDOPATH*5MM CURVED SCISSORS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 5DCS
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the facility received boxes of product for a trial and upon using the product they would pop open.No patient harm was reported.
 
Additional Manufacturer Narrative
(b)(4).Date sent 7/9/2025.D4 batch #a97h1e.B3: only event year known: 2025.Investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the 5dcs device was returned with no damage in the external components.The instrument was mechanically tested in order to evaluate the reported incident and during the analysis, it was noted that the tip of the jaws was misalignment causing the difficulty to close and open the trigger.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of ethicon¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event.The reported complaint was confirmed.A manufacturing record evaluation was performed for the finished device batch a97h1e number, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: what is meant by "the product would pop open"? was there any damage to the device? was there any damage to the jaw? was there any damage to the handle? if other, please specify.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 12/29/2025.D4: batch # a97h1e.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the 5dcs device was returned with no damage in the external components.The instrument was mechanically tested in order to evaluate the reported incident and during the analysis, it was noted that the tip of the jaws was misalignment causing the difficulty to close and open the trigger.As part of ethicon¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event.The reported complaint was confirmed.Device history review a manufacturing record evaluation was performed for the finished device batch a97h1e number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
3035526892
MDR Report Key22456861
Report Number3005075853-2025-05298
Device Sequence Number18215556
Product Code GEI
UDI-Device Identifier10705036000785
UDI-Public(01)10705036000785(17)300131(10)A97H1E
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K984240
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5DCS
Device Lot NumberA97H1E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/08/2025
Supplement Date Received by Manufacturer12/04/2025
Initial Report FDA Received Date07/09/2025
Supplement Report FDA Received Date12/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-