• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FERNO-WASHINGTON, INC. PRO 28Z CHAIR COT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FERNO-WASHINGTON, INC. PRO 28Z CHAIR COT Back to Search Results
Model Number 0012826
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/19/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the diagonal leg brace was observed to be broken during a patient transfer and the legs of the stretcher would not retract into the stretcher for loading into the ambulance.It was later confirmed by the end user that they were able to manually manipulate the legs and the stretcher was able to be utilized to complete the patient transfer and there were no injuries or adverse effects as a result of the product issue.An authorized field technician was dispatched to conduct a visual and functional evaluation of the stretcher at the customer's location.The reported issue was confirmed and the stretcher was repaired and returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRO 28Z CHAIR COT
Common Device Name
PRO 28Z CHAIR COT
Manufacturer (Section D)
FERNO-WASHINGTON, INC.
70 wiel way
wilmington OH 45177
Manufacturer (Section G)
FERNO-WASHINGTON, INC.
70 wiel way
wilmington OH 45177
Manufacturer Contact
dawn greene
70 wiel way
wilmington, OH 45177
9372832900
MDR Report Key22457128
Report Number1523574-2025-00072
Device Sequence Number8118327
Product Code FPO
UDI-Device Identifier00190790002671
UDI-Public(01)00190790002671(11)230321
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0012826
Device Catalogue Number0012826
Device Lot NumberM562090
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2025
Initial Report FDA Received Date07/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/21/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexMale
Patient Weight68 KG
-
-