| Catalog Number |
0250070500 |
| Medical Device Problem Codes |
Leak/Splash (1354); Overheating of Device (1437); Smoking (1585)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the suction irrigator for was being used for 15-30 minutes, the tip started leaking.Staff attempted to unstick button, in case that was causing it to run and dispense saline outside patient, did not fix the issue.Surgeon observed that the tip was "smoking." at this time, suction irrigator was removed from sterile field, replaced with new device.Motor for irrigator was still running at this time, as if the button was fully depressed, battery compartment/motor where irrigator spikes saline bag hot to the touch.Therefore, there was a thermal event.
|
| |
|
Event or Problem Description
|
|
It was reported that the suction irrigator for was being used for 15-30 minutes, the tip started leaking.Staff attempted to unstick button, in case that was causing it to run and dispense saline outside patient, did not fix the issue.Surgeon observed that the tip was "smoking." at this time, suction irrigator was removed from sterile field, replaced with new device.Motor for irrigator was still running at this time, as if the button was fully depressed, battery compartment/motor where irrigator spikes saline bag hot to the touch.Therefore, there was a thermal event.
|
| |
|
Additional Manufacturer Narrative
|
|
Three attempts have been made to retrieve the device for evaluation, and it is yet to be received in-house.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure : tip was smoking probable root cause for equipment leaks and flow too high : 1.Material/design error 2.Constant irrigation flow is not maintained leading to limited field of view 3.Stopcocks in improper configuration 4.Large anatomy 5.Missing o-ring 6.Damaged or deformed o-ring 7.Pump malfunction (motor, control board, harness, power supply, battery, or cassette) 8.Clogged tubing 9.User error probable root cause for smoke smell or flames coming from device : 1.Insulation melting 2.Excessive cauterization 3.User error 4.Nonconforming electrical components probable root cause for overheating: 1.Material/design error 2.User error the reported failure mode will be monitored for future reoccurrence.
|
| |
|